Veris-technologies Quad 1000 Soil EC Mapping System - Operating Instr Manual de usuario Pagina 24

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 43
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 23
OM17-Quad1000
6-3
Figure 2.2
Firmly press one lead of the ohmmeter to the #1 disk blade edge (left hand, standing behind the
unit) and the other lead to the terminal marked #1on the Quad 1000 test box decal. Maintain firm
pressure on the ohmmeter lead touching the disk blade. A reading of less than 2 ohms is normal.
Rotate blade ¼ of a turn back and forth as you view the ohmmeter. Any jump in the readings
above 2 ohms indicates a problem. (Figure 2.3)
Figure 2.3 Testing signal continuity on Quad 1000 #1 disk.
Continue to check each disk electrode in succession, left to right.
If any disk electrode exhibits no continuity or resistance higher than 2 ohms, refer to Procedure #3
Diagnosing EC Signal Problems.
Connect other lead
to disk blade
Ohmmeter
Connect one lead to Signal Test
Box terminal corresponding to
disk blade
Signal extension cable
(from implement)
Vista de pagina 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 42 43

Comentarios a estos manuales

Sin comentarios